Laboratory Gas Burner

Instructions for cylinder exchange

It is mandatory to handle the device in a chemical hood or in a well ventilated area, away from an open flame or sparks

No. Picture Instruction
Instructions for cylinder exchange
1 Instructions for cylinder exchange Close the gas valve. Make sure that the used cylinder is empty by shaking it to check that no liquid can be heard inside.
2 Instructions for cylinder exchange Unscrew the gas valve unit completely.
3 Instructions for cylinder exchange Remove the bottom of cylinder housing. Bring the reed to the aperture of the burner and remove the bottom.
4 Instructions for cylinder exchange Remove the empty cylinder and keep overnight for fully vaporation.
5 Instructions for cylinder exchange This is how the installation should look before replacing a new cylinder.
6 Instructions for cylinder exchange This is how the installation should look before replacing a new cylinder.
7 Instructions for cylinder exchange Align and insert a new cylinder into the cylinder housing.
8 Instructions for cylinder exchange Install the bottom of the cylinder housing. Make sure that the bulges are directed towards the cylinder. Turn the bottom.
9 Instructions for cylinder exchange Make sure that the needle housing is indeed spring-like and that the needle does not protrude. Make sure that the valve is closed.
10 Instructions for cylinder exchange Screw the gas valve unit into the cylinder housing until it is fastened tightly.