The Weizmann institute created a human sample bank allowing Weizmann scientiscts to store sample originated from human subjects in accordance with the bioethics regulation. All samples are stored in a fully automatic -80 storage by LiCONiC. The system is designed to store different Human sample types (tumor, blood, DNA etc.) The robotic freezer system is designed to maximize sample preservations at all times. Temperature is preserved by a main compressor and has 2 additional backups and alarming system to notify temperature changes at any time. The -80 chamber is completely isolated from room temperature and humidity. Sample handling (barcode scanning, tube picking) in the interface unit at -20.
- The biobank is equipped with 1D and 2D barcode scanners for scanning suitable racks and tubes.
- In addition a controlled rate freezer (ice cube by SY-LAB) is available to increase viability for cell line and tissue samples.
System capacity
Total tube capacity - 288,096
Total racks - 3412
- 0.3 ml - 1920 racks (96 tubes/rack by LVL)
- 1 ml - 580 racks (96 tubes/rack by LVL)
- 2 ml - 912 racks (48 tubes/rack by LVL)
- 4 ml - 231 racks (48 tubes/rack by Micronic)
- 4.5 ml - 468 racks (48 tubes/rack by LVL)
Sample management: RBF software
The robotic system is operated by a Weizmann
developed sample handling system. (manages
sample data and controls the RBF operation)